logo SBA

ETD

Archivio digitale delle tesi discusse presso l’Università di Pisa

Tesi etd-01282025-102254


Tipo di tesi
Tesi di laurea magistrale
Autore
MOZZO, CATERINA
URN
etd-01282025-102254
Titolo
Plastic-scintillator based dosimetry for minibeam radiotherapy
Dipartimento
FISICA
Corso di studi
FISICA
Relatori
relatore Dott.ssa Ciarrocchi, Esther
Parole chiave
  • dosimetry
  • minibeam radiotherapy
  • optical imaging
  • plastic scintillator
Data inizio appello
17/02/2025
Consultabilità
Non consultabile
Data di rilascio
17/02/2028
Riassunto
Minibeam radiotherapy (MBRT) is a dose delivery technique that exploits the alternation
of high-dose ("peaks") and low-dose ("valleys") regions, with the potential goal of achieving
a higher sparing of healthy tissues without compromising tumor control compared to
conventional radiotherapy. The involved beamlets are millimetric, and single temporal fractions
are used. Ultra High Dose Rate (UHDR) radiotherapy is another modality that offers
the potential to achieve these outcomes by eliciting the so-called FLASH effect, a peculiar
tissue response generated by ultra-fast beam delivery at average dose rates thousands of
times larger than conventional radiotherapy. Both irradiation methods present significant
challenges from a dosimetric perspective: UHDR beams often induce a non-linear response
in dosimeters, due to the high deposited dose in a brief time interval, and MBRT requires
fine spatial resolution to resolve the beamlet structures. Plastic-scintillator based dosimeters
may be suitable candidates for both irradiation modalities, given their potentially high spatial
resolution and linearity with dose and dose rate, besides being almost tissue-equivalent
and more cost-effective than other detector technologies.
The main objective of this thesis was to assess the dosimetric performance of a plastic scintillator
based detector for low-energy electron FLASH beam and minibeam radiotherapy.
The developed dosimeter consisted of a 0.5-mm thick sheet of EJ212 plastic scintillator
placed in a custom PVC holder and irradiated with 9-MeV electrons, generated by the
Triode-Gun equipped ElectronFlash LINAC, available at the Centro Pisano for FLASH Radiotherapy
(CPFR). A 45◦ inclined mirror was used to reflect the scintillation light emitted
by the plastic sheet toward a scientific CCD camera coupled with an optical lens. The
performance of the dosimeter was evaluated by assessing the dose-response linearity, the
capability to measure the percentage depth dose in solid water, the spectrum of the plastic
scintillator emission, the ability to measure minibeam dosimetric parameters, and the spatial
resolution of the setup.
The linearity was tested by varying the beam Dose Per Pulse (DPP). The percentage depth
dose curve was measured at a fixed output dose by adding layers of solid water with increasing
thickness before the scintillator. In this case, the images were acquired both unfiltered
and with three optical filters (centered at 405 nm, 457 nm, and 532 nm) to assess the
Cerenkov radiation contribution and correct the PDD to obtain a better agreement with the
reference curve (measured with a calibrated FLASH diamond detector). The EJ212 spectral
emission was analyzed with a spectrograph and fiber-optic system. The electron beam was
placed at various depths in solid water, and a sufficiently low-energy X-ray beam was used
to measure the scintillation spectrum alone without Cerenkov emission. The capability to
image spatially-fractionated dose patterns was investigated, with different minibeam collimators,
varying the DPP, and the depths in solid water. Three parameters were evaluated:
the peak-valley dose ratio (PVDR), the mean center-to-center (CTC) distance of adjacent
peaks, and the peak full-width-half-maximum (FWHM). Finally, the spatial resolution of
the scintillator and imaging system was evaluated using an X-ray source and images of an
edge. The contribution of the imaging system alone (with no scintillator) was estimated
using optical patterns.
Linearity was found to be verified up to 12 Gy per pulse, corresponding to 3 MGy/s of intra-pulse dose rate. The PDD was correctly measured with the dosimeter. The Cerenkov
correction enabled a better agreement with the reference PDD, reducing the residuals from
6% to 3% vs. the reference. All the spectra were mainly coherent with the tabulated EJ212
emission (peak wavelength at 423 nm). A mismatch with the tabulated spectrum was found
around 405 nm, presumably due to thickness-dependent self-absorption effects. Moreover,
the Cerenkov light contribution could be observed above 500 nm in the spectrum obtained
with electrons. The spatial features of fractionated beams (CTC and FWHM) were correctly
reconstructed within ∼ 6% of the reference values. For the PVDR values, since no
reference simultaneous measurements were acquired, the only comparison could be made
with a recently published paper, which seems to suggest slightly larger values than the ones
measured in this work. This difference could be partially due to current limits in the spatial
resolution of the imaging system and to slight differences in the data analysis. However,
the main contributor is presumably a technical intervention on the LINAC between the two
measurements, which affected the beam energy spectrum and the delivered dose. The spatial
resolution was found to be in the tenths of a millimeter range.
The results suggest that this setup based on plastic scintillator sheets is appropriate for
low-energy electron FLASH and MBRT dosimetry applications. Future improvements will
be related to improving the spatial resolution, validating the measurements of dose spatial
fractionation with a reference detector, and better correcting the Cerenkov contribution.
File